User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

Agilent B1500 User’s Guide, Edition 7 4- 47
Using EasyEXPERT
Configuration
Cable Specifies the model number of the cable from the
B2200A/ B2201A output to the DUT I/F. The selections
are the 16494A/B/C-001 (1.5 m) / 002 (3 m) / 005 (4 m).
This mode cannot perform the capacitance compensation fo
r the cables to
positioner/manipulator or the probe card.
• User Compensation Data File
Select this radio button and specify the compensation data file to perform the
capacitance compensa
tion for the path from the B2200A/B2201A input
ports to the end of positioner/manipulator/probe card. See “To Create
Compensation Data File” on page 3-50. The data after compensation is
displayed and saved.
• E5250A (option)
The following setup is available for the E5250A hardware profile.
• User Compensation Data File
Select this radio button and specify the compensation data file to perform the
capacitance comp
ensation for the path from the E5250A input ports to the
end of positioner/manipulator/probe card. See “To Create Compensation
Data File” on page 3-50. The data after compensation is displayed and
saved.
Clear Clears the User Compensation Data File field.
Triaxial Cable Enter the length of the 16494A triaxial cable
connected between the E5
250A outputs to the
connector plate.
User Triaxial Cable Enter the length of the triaxial cable between
the connector plate
and the DUT. Enter 0 if
the triaxial cable is not used.
User Coaxial Cable Enter the length of the coaxial cable between
the connector plate
and the DUT. Enter 0 if
the coaxial cable is not used.
Close Closes the Extended Configuration dialog box.