User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

Agilent B1500 User’s Guide, Edition 7 4- 45
Using EasyEXPERT
Configuration
Extended
Configuration
This button opens the “Extended Configuration” on page 4-46 dialog box.
SMU/PG Selector
The SMU/PG Selector tab screen is used to specify the input connection and the
channel connection status of the Agilent 16440A SMU/PG selector. The selector can
be controlled by the B1500A via the Digital I/O connector and the Agilent 16445A
selector adapter.
Two selectors can be used. The first selector must be
connected to the Digital I/O
connector via the 16445A. And if the second selector is used, it must be connected
to the first selector’s Control Output connector. See “SMU/PG Selector” on page
7-48.
Enable SMU/PG
Se
lector
To
use the selector, connect the selector to the B1500A, and check this check box.
Selector CH1/CH2 CH1 and CH2 indicates the channels 1 and 2 of the first selector respectively.
Selector CH3/CH4 CH3 and CH4 indicates the channels 1 and 2 of the second selector respectively.
SMU Assignment Specifies the SMU connected t
o the SMU inpu
t of the selector channel CH1, CH2,
CH3, or CH4.
Input/Output Path Specifies the connection status of the selector channel CH1, CH2, CH3, or CH4.
If this field is set to Always SMU, the selector makes the path from Input SMU to
Output.
If th
is field is set to Normally PGU (AUX), the selector performs automatic
switching in every
test. The selector normally makes the path from Input PGU to
Output and makes the path from Input SMU to Output only for the test which uses
the SMU connected to the Input SMU terminal.
The CH1 and CH3 also provides the PGU Open status which is set by using the
Advanced window
of the Measurement Setup.
Event Log
The event log tab screen displays the event log.
Date/Time Displays date and time of the event.
Message Event message