User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

Agilent B1500 User’s Guide, Edition 7 4- 43
Using EasyEXPERT
Configuration
ASU
The ASU tab screen is available if the B1500A installs the HRSMU calibrated with
the ASU (Atto Sense/Switch Unit) regardless of the ASU connection status, and
shows the ASU information.
Slot Slot number
Module Type Type of module
Name Name of module
ASU Serial
Number
Serial number of the usable ASU. This cell shows the serial number even if the ASU
i
s disconn
ected.
If an incorrect ASU is connected to the HRSMU, this cell di
splays an error message
instead of the serial number. Error message is also displayed if an ASU is connected
to the HRSMU which is not available for using ASU.
ASU I/O Path ASU input to output connection, SMU or AUX
If this field is set to SMU, the ASU makes the path from the ASU Force input to the
ASU output.
If
this field is set to AUX, the ASU performs automatic switching in every test. The
ASU normally make
s the path from the AUX In input to the ASU output and makes
the path from the ASU Force input to the ASU output for the test with the HRSMU.
Switching Matrix
The Switching Matrix tab screen is used to confirm the GPIB connection with
Agilent B2200A/B2201A/E5250A switching matrix, and defines the connections
between the measurement units and the switching matrix input ports.
Switching Matrix
Ty
pe
Displays th
e type of the switching matrix. One of the following.
• B2200A Femto Leakage Switch Mainframe (standard)
• B2201A 14ch Low Leakage Switch Mainframe (standard)
• E5250A Low Leakage Switch Mainframe (option)
• No Switching Matrix (if switching matrix is not connected)
GPIB Address Specifies the GPIB address of the switching m
atrix. 1 to 30.