User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

3- 24 Agilent B1500 User’s Guide, Edition 7
Installation
Installing Accessories
2. Fix the SCUU inside or outside the shielding box. The SCUU must be fixed to
the best position for accessing its connectors. Agilent N1301A-110 Magnetic
Stand will be useful for fixing the SCUU.
3. Attach the other end of the SCUU cable to the B1500A
rear panel connectors for
the MFCMU in the slot N and two SMUs in the slots N-1 and N-2 (N: integer, 3
to 10).
Connection will be easy by bringing the control connectors together before
bringing the B
NC connectors.
4. Fix the GSWU in the shielding box. The GSWU must be fixed to the best
position fo
r accessing its connectors.
5. Connect the GSWU cable (Agilent N1301A-201/202) between the SCUU and
the GSWU.
If th
e SCUU is fixed outside the shielding box, pass the GSWU cable through
the cable ho
le of the 16495K.
6. Adjust the cable length in the shielding box, and catch in the cables by using the
cover of the 164
95K.
7. Connect the proper cables from Force1/CMUH and Fo
rce2/CMUL to the 16495
connector plates or the DUT interface. See Figure 3-5.
For the Kelvin connections, also connect the proper cables f
rom Sense1 and
Sense2 to the 16495 connector plates or the DUT interface.
8. Connect the wires (furnished, 2 ea.) to the GSWU.
9. Connect one of the wires to the Guard terminal exten
ded from CMUH of SCUU.
Then the wire should contact the Guard terminal as close to the device as
possible.
10. Connect another one to the Guard terminal extended from CMUL of SCUU.
Then the wire
should contact the Guard terminal as close to the device as
possible.
NOTE About SCUU output cables
To perform capacitance measurements accurately, the cable length between DUT
and the SCUU output must be as short as possible.