User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

3- 18 Agilent B1500 User’s Guide, Edition 7
Installation
Installing Accessories
To install ASU 1. Fix the ASU in the shielding box. The ASU must be fixed to the best position for
accessing its connectors. Agilent N1254A-108 Magnetic Stand will be useful for
fixing the ASU.
2. Pass the 16493M (D-sub cable, triaxial cable), N1300A CMU cable, 16048D/E
test leads, BNC cable an
d so on through the cable hole of the 16495K. They will
be connected between the ASU and the instruments
3. Connect the D-sub cable to the ASU D-sub connector.
Connect the triaxial cable to the ASU Force terminal.
Connect the cable to the ASU CMU-cur/AUX In terminal. See Figure 3-2.
• For the MFCMU, connect the N1300A CMU cable.
• For Agilent 4284A LCR meter, connect Agilent 16048D/E test leads.
• For the instrument other than the MFC
MU or the 4284A, con
nect the BNC
cable to the CMU-cur/AUX In terminal. And open the CMU-pot terminal or
cover it by using the BNC open cap.
4. Adjust the cable length in the shielding box, and catch in the cables by using the
cover of the 164
95K.
5. Connect the cable extended from the DUT interface to the ASU output terminal.
Fo
r the non Kelvin connection, open the Sense terminal or cover it by using the
triaxial open cap.
T
o connect
B15
00A
1. Turn t
he B1500A off.
2. Connect the D-sub cable from the ASU to the HRSMU D-su
b connector.
3. Connect the triaxial cable from the ASU to the HRSMU Force terminal.
NOTE The specifications of the ASU are guaranteed while it is connected to the exclusive
HRSMU. Confirm the serial number of the ASU, and connect it properly. See
“ASU” on page 4-43.
To connect other
instrument to AUX
To connect the instrument other than the MFCMU or the 4284A:
1. Turn the instrument off.
2. Connect the BNC cable from the ASU CMU-cur/AUX In
terminal to the
instrument. Prepare an adapter if the instrument’s input/output connector is not
BNC.