User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

3- 16 Agilent B1500 User’s Guide, Edition 7
Installation
Installing Accessories
Connecting the interlock terminal
The B1500A provides the Interlock connector to prevent you from receiving an
electrical shock from high voltage (more than ± 42 V). If the Interlock connector is
open, the B
1500A cannot force high voltage more than ± 42 V.
When using the 16442B test fixture, connect the Interlock connectors together using
the 16493J i
nterlock cable.
If you use the 16495F/G/H/J connector plate, your own connector plate, or your own
test fi
xture, make an interlock circuit as shown in “To Make an Interlock Circuit” on
page 3-29. Then connect the B1500A Interlock connector to the external inter
lock
connector by using the 16493J interlock cable.
If you use a connector plate or test fixture that has a BNC coaxial connector for the
interlock connecto
r, connect the BNC interlock connector to the B1500A interlock
terminal as follows:
• Required Parts:
• Procedure:
1. Connect the Interlock connector on the B1500A rear panel to the interlock
cab
le adapter us
ing the 16493J interlock cable.
2. Connect the interlock cable adapter to the BNC i
nterlock connector using the
coaxial cable furnished with the 16435A.
Model No. Quantity Description
Agilent 16435A 1 Interlock Cable Adapter
Agilent 16493J 1 Interlock Cable