User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

Agilent B1500 User’s Guide, Edition 7 3- 11
Installation
Installing Accessories
Installing Accessories
This section describes how to install the B1500A and accessories. Additional
information regarding airflow can be found in the “Ventilation Requirements” on
page 3-4.
This section describes how to:
• “To Connect 16442B Test Fixture”
• “To Connect Connector Plate”
• “To Connect ASU”
• “To Connect SCUU/GSWU”
• “To Connect GNDU Adapter”
• “To Interconnect SPGUs”
NOTE Before connecting accessory to MFCMU
Turn the B1500A off before connecting the SCUU, the SCUU cable, or the CMU
cable to the MFCMU.
WARNING There are potentially hazardous voltages of up to ± 100 V (MPSMU/HRSMU)
or ± 200 V (HPSMU) at the Force, Sense, and Guard terminals. To prevent
electrical shock, do not expose these lines.
WARNING To prevent electrical shock during use, install an interlock circuit (see
page 3-29) and connect the Interlock cable (see page 3-16).
CAUTION The GNDU can sink current of up to 4.2 A. Use Agilent 16493L GNDU cable to
connect the GNDU to a test fixture or a connector plate.
Do not use normal triaxial cable (Agilent 16494A) because the maximum current
rating of the cable is 1 A.
CAUTION Never connect the Guard terminal to any output, including circuit common, chassis
ground, or any other guard terminal. Doing so will damage the SMU.