User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

2- 68 Agilent B1500 User’s Guide, Edition 7
Introduction
Specifications
Measurement time resolution: 100 μs
• Standby mode
SMUs in “Standby” remain programmed to their specified o
utput value even as
other units are reset for the next measurement.
• Current offset cancel
This function subtracts the offset current from the current measurement raw
data, an
d returns the result as the measurement data. This function is used to
compensate the error factor (offset current) caused by the measurement path
such as the measurement cables, manipulators, or probe card.
• Timestamp
The B1500A supports a time stamp function utilizing an internal quartz clock.
Resolution: 100 μs
• Other measurement characteristics
• Measurement control:
Single, repeat, append, and stop
• SMU setting capabilities:
Limited auto ranging, voltage/current compliance, power compliance,
au
tomatic s
weep abort functions, self-test, and self-calibration
Arithmetic and analysis functions
• User functions
Up to 20 user-defined functions can be defined using arithmetic expressions.
Measured data and pre-defined variables can be used in the computation. The
results can be di
splayed on the LCD.
• Arithmetic operators
+, -, *, /, ^, abs (absolute value), at (arc tangent), avg (averaging), cond
(conditio
nal evaluation), delta, diff (differential), exp (exponent), integ
(integration), lgt (logarithm, base 10), log (logarithm, base e), mavg (moving
average), max, min, sqrt, trigonometric function, inverse trigonometric function,
and so on
• Physical constants
Keyboard constants are stored in memory as follows: