Lab Exercise: Advanced Reporting
Automated Spectral Library Search
131
to the calibration table for an identically named entry. If one is found, the data are
used to calculate and report the amount.
6) OK this dialog box. Open the library you intend to use in the report and
check that the names of the calibrated peaks are identical with the library
names. In other words, check your calibration table as well as the entries in
the library.
7) Under Specify Report, select Library Search. OK the dialog box. Resave
your method. Print the report.
8) Now, set up a method where the search mode is Target compound analysis
using a calibration table.
This search mode is only for those compounds in the calibration table. Peaks are
only reported as identified if confirmed by spectral comparison.
9) Print the report.