Troubleshooting guide
Lab Exercise: Quantification
Setting up Low and High Amount Limits for Calibration Standards
96
Setting up Low and High Amount Limits for
Calibration Standards
You may specify low and high amount limits in the calibration table. If the
calculated amount is outside the limits, a warning appears in the report. The
amount limits can be used as qualifiers to confirm the peak identification.
Compounds that are outside of the amount limits, are not used to recalibrate the
Calibration Table. The minimum and maximum amounts limits do not consider a
multiplier value that you may have specified.
1) From the Calibration menu, select Calibration Table Options..., then
Compound Details or change from Overview to Compound Details in the
drop box located in the middle of the window.
2) Click in the box presented to the left of the Low Limit column while in the
biphenyl row. Now click in the Low Limit cell. A lower limit is
automatically calculated. This is a lower limit of -20%.
3) Click in the box presented to the left of the High Limit column while in the
biphenyl row. An upper limit will appear in the cell. This is an upper limit of
+20%. The default is +/- 20%, but you may set any desired limits.
4) Repeat this procedure for the o-terphenyl peak.
5) Resave your method.