Technical data

2 Calibration
10 U2761A Service Guide
Introduction
Self-Test
A brief power-on self-test occurs automatically whenever the
U2761A is turned on. This limited test assures that the
instrument is capable of operation. To perform a complete
self-test, send the following SCPI command to the U2761A.
*TST?
The U2761A will automatically perform the complete self-test
procedure when the SCPI command is sent. The self-test will
be completed in a few seconds.
If the self-test is successful, a zero (0) is returned.
If the self-test fails, an error number is returned.
Agilent Measurement Manager
To start the self-test procedure using the Agilent
Measurement Manager, perform the following test.
1 Click Tools > Self-Test.