Output Configuration Commands 4
U2761A Programmer’s Reference 49
[SOURce:]VOLTage[:LEVel][:IMMediate]:LOW
Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate]:LOW <voltage>
This command sets the low voltage level.
[SOURce:]VOLTage[:LEVel][:IMMediate]:LOW?
This query returns the low voltage level for the currently selected
function.
Parameter
Examples
The following command sets the low voltage level to –2 V.
VOLT:LOW –2
The following query returns the low voltage level.
VOLT:LOW?
Typical response:
–2.0000000000000E+00
Item Type Range of values Default value
voltage Numeric –2.5 V to 2.46 V (Into 50 Ω load)
–5 V to 4.92 V (Into open circuit)
–0.5 V
SOUR : VOLT space
?
voltage:LEV :IMM :LOW