Technical data
4 Output Configuration Commands
42 U2761A Programmer’s Reference
[SOURce:]FUNCtion[:SHAPe]
Syntax
[SOURce:]FUNCtion[:SHAPe] {SINusoid|SQUare|RAMP|PULSe|DC|
USER}
This command sets the output function.
[SOURce:]FUNCtion[:SHAPe]?
This query returns the selected output function as SIN, SQU, RAMP, PULS,
DC, or USER.
Parameter
Remarks
For the USER output function, you will need to download user- defined
Arbitrary waveform data into the U2761A volatile memory. Refer to
Chapter 10, “Arbitrary Waveform Commands” on page 103 for more
information on downloading Arbitrary waveforms to memory.
Item Type Range of values Default value
function Discrete SINusoid|SQUare|RAMP|PULSe|DC|USER SINusoid
SOUR : FUNC space:SHAP SIN
SQU
RAMP
PULS
DC
USER
?