Technical data
16 Error Messages
154 U2761A Programmer’s Reference
708 Calibration error; perform setting is not allowed
The calibration setting was predetermined by calibration setup and modification to the setting is not
allowed
755 Calibration error; timebase calibration is not completed
The timebase calibration steps have not fully completed before exiting calibration mode
756 Calibration error; DC offset calibration is not completed
The DC offset calibration steps have not fully completed before exiting calibration mode
757 Calibration error; AC amplitude calibration is not completed
The AC amplitude calibration steps have not fully completed before exiting calibration mode
758 Calibration error; 0 dB flatness calibration is not completed
The 0 dB flatness calibration steps have not fully completed before exiting calibration mode
759 Calibration error; –20 dB flatness calibration is not completed
The –20 dB flatness calibration steps have not fully completed before exiting calibration mode
760 Calibration error; +20 dB flatness calibration is not completed
The +20 dB calibration steps have not fully completed before exiting calibration mode
761 Calibration error; output impedance calibration is not completed
The output impedance calibration steps have not fully completed before exiting calibration mode
850 Calibration error; setup is invalid
You have specified an invalid calibration setup number with the CAL:SET command
851 Calibration error; setup is out of order
Certain calibration setups must be performed in a specific sequence in order to be valid
852 Calibration error; setup timebase calibration is not in sequence
The setup timebase calibration steps are not in sequence as required
Tabl e 16-7 Calibration errors (continued)