Technical data
Error Messages 16
U2761A Programmer’s Reference 153
Self-test errors
The table below shows the self- test errors.
Calibration errors
The calibration errors are listed in the following table.
Tabl e 16-6 Self-test errors
Error number Self-test errors
604 Self-test failed; waveform memory bank
Either the waveform RAM or the synthesis IC has failed
605 Self-test failed; modulation memory bank
The modulation memory bank in the synthesis IC has failed
630 Self-test failed; internal ADC over-range condition
Indicates a probable ADC failure which could be of the system ADC, ADC input multiplexer, or ADC input
buffer amplifier
631 Self-test failed; internal ADC measurement error
Indicates a probable ADC failure which could be of the system ADC, ADC input multiplexer, or ADC input
buffer amplifier
Tabl e 16-7 Calibration errors
Error number Calibration errors
702 Calibration error; calibration memory is secured
A calibration cannot be performed when calibration memory is secured
703 Calibration error; secure code provided was invalid
Invalid security code specified with the CAL:SEC:STAT ON command
706 Calibration error; provided value is out of range
The calibration value specified with the CAL:VAL command is out of range