Technical data

Error Messages 16
U2761A Programmer’s Reference 151
Device-dependent errors
The table below contains the device-dependent errors.
–222 Data out of range; phase offset; value clipped to lower limit
The phase offset is limited to the lower limit by the U2761A hardware
–222 Data out of range; phase offset; value clipped to upper limit
The phase offset is limited to the upper limit by the U2761A hardware
Tabl e 16-2 Execution errors (continued)
Tabl e 16-3 Device-dependent errors
Error number Device-dependent errors
–310 System error
The U2761A operation has not completed properly, possibly due to an abnormal hardware or firmware
condition
–313 Calibration memory lost; memory corruption detected
The nonvolatile memory used to store the U2761A calibration constants has detected a checksum error
which may be caused by a device failure or extreme conditions such as lightning or strong magnetic fields
–315 Configuration memory lost; memory corruption detected
The nonvolatile memory used to store the U2761A configuration settings has detected a checksum error
which may be caused by a device failure or extreme conditions such as lightning or strong magnetic fields
–350 Queue overflow
The error queue is full because more than 20 errors have occurred. No additional errors are stored until you
remove the errors from the queue.