User's Manual
 Getting Started
38
Voltage Test
 1. Set the voltage of the selected output to 10 V by pressing:
 VSET  1  0  ENTER
 2. Check that the display reads approximately 10 V and 0 A and the CV annunciator is on indicating that the supply is in
the constant voltage mode of operation.
Overvoltage Test
 1. Program the overvoltage protection (OVP) to 19 V by pressing:
 OV  1  9  ENTER
 SET    
 2. Set the voltage to 16 V by pressing:
 VSET  1  6  ENTER
 3. Check that the display reads approximately 16 V and 0 A.
 4. Set the voltage to 20 V by pressing:
 VSET  2  0  ENTER
 5. Check that the display reads "OVERVOLTAGE’’
 6. Reset the supply by pressing:
 VSET  1  6  ENTER  RESET
 7. Check that the display reads approximately 16 V and 0 A.
Current Test
 1. Turn off the supply.
 2. Remove the barrier block cover from the output to be tested and connect a short circuit (jumper wire) between the + V
and -V terminals of the output being tested.
 3. Turn on the supply.
 4. Use the 
 to select the output being tested.
 5. Set the voltage to a volts by pressing:
 VSET  5  ENTER
  Set the current to 0.01 amps by pressing:
 ISET  .  0  1  ENTER