User's Manual
 Remote Operation
 71
Initial Conditions
 Immediately after power on from the factory, the power supply automatically undergoes a self-test and sets all parameters
to the values contained in Table 5-3. The values in the first part of the table come from storage register 0 and were stored at
the factory. They will remain until STO 0 is used to change settings of the power on state. When the power supply is again
turned on, it will initialize its outputs to the values in storage register 0.
 Table 5-3. Initial Conditions
 Parameter  Inital Value
 Voltage  0 V High Range
 Current  10 mA High Range
 Reprogramming Delay  20 mS
 Store/Recall registers:  
  0-10: Voltage  0 V High Range
  Current  10 mA High Range
 Over Voltage (OV)  55 V (High Range)
 Over Current Protection (OCP)  Disabled
 UNMASK register
 0 (cleared)
 SRQ  0 (Off)
 Output Channels  On
 Front Panel Metering  Output #1
 Power Supply Address  Last stored value (Factory set to 5)
 Local Control  Enabled
 PON bit  On
 PON SRQ  Off (0)
 Cal Mode  Off
Power Supply Commands
 This section discusses the commands which you will use to program the supply’s voltage and current, protection circuits,
and enhanced features like storage and recall registers, and reprogramming delay. When programming, you should be
aware that the current, voltage and overvoltage ranges for each output of your supply may differ. Table 5-1 shows these
values for the power supply. If you send values out of these ranges, you will get a number range error. A summary of all
commands appears in Appendix C.
 The output voltage of some output channels exceeds the safe operating limit of 42.2 V. To avoid any
electrical shock, program the voltage to zero volts or turn off ac input power before changing any rear
panel connections. Make certain all straps are properly connected, terminal block screws are securely
    tightened and terminal block covers are replaced before reapplying power.
Voltage Programming
 To program voltage, send the output channel and the programmed value. In the example below, output l is programmed to 5
V.
 VSET 1,5
 The values you send must always be volts. For example if you want to program 450 millivolts, convert to volts and then
send the command:










