User's Manual
82 
Programming Reference
SCPI Instrument Command List
Command
:STATus:PRESet
Long
:STATus:PRESet
Form
Event
*RST value
Not Applicable
Description
This command
 
Clears all status group event-registers
 
Clears the error queue
 
Presets the status group enable-, PTR-, and NTR-registers as follows:
Command
:STATus:QUEStionable
This command tree accesses the QUEStionable status group. The 
QUEStionable status group contains warning bits for voltage, current, 
time and frequency parameters. A warning occurs when the output signal 
could be out of specification due to the combined specification 
uncertainties of many parameters, although all parameters are set within 
their individually specified limits. If a parameter is set outside its 
specified limits an error is generated.
The following commands are used to access the registers within the 
status group:
Status Group Register Preset value
OPERation ENABle
0000000000000000
PTR
0111111111111111
NTR
0000000000000000
QUEStionable ENABle
0000000000000000
PTR
0111111111111111
NTR
0000000000000000