User's Manual
78 
Programming Reference
SCPI Instrument Command List
Specified Limits
Agilent 81131A: 1.95 V to 3.75 V
Agilent 81132A: 1.95 V to 2.95 V
Value coupling
Range coupling
Amplitude
Description
This command programs the offset voltage of the OUTPUT signal. Note 
that to set the OUTPUT levels in terms of voltage, you first have to 
execute the 
[:SOURce]:HOLD VOLTage
 command to enable the 
[:SOURce]:VOLtage
subsystem.
The available voltage range is limited by the specified current limits.
Example
:HOLD VOLT
Enable VOLTAGE subsystem
:VOLT1:OFF -800MV
Set OUTPUT 1 offset
to 800mV
Command
:VOLT[1|2]:HIGH
Long
[:SOURce]:VOLTage[1|2][:LEVel][:IMMediate]:HIGH
Form
Set & Query
Parameter
Numeric
Parameter suffix
V
 with engineering prefixes.
*RST value
500 mV
Specified Limits
Agilent 81131A: 1.90 V to 3.80 V
Agilent 81132A: 1.90 V to 2.50 V
Value coupling
High = Offset + 
Amplitude
2
Low = Offset – 
2
Amplitude
Amplitude = High – Low
 Offset = 
High – Low
2