User's Manual
 45
Programming Reference
SCPI Instrument Command List
Command
:INIT:CONT
Long
:INITiate:CONTinuous
Form
Set & Query
Parameter
ON | OFF | 1 | 0
*RST value
ON
Description
Use this command to enable/disable automatic restart of the instrument 
(equal to start and stop the instrument). If 
:ARM:SOURce
 is set to 
IMMediate
, the value of 
:INITiate:CONTinuous
 is ignored.
Command
:CHAN:MATH
Long
:CHANnel:MATH
Form
Set & Query
Parameter
OFF | DIGital 
*RST value
OFF
Description
Use this command to enable or disable digital channel addition in an 
instrument with two Output channels installed. 
With 
:CHAN:MATH DIGital
 the digital signals from both channels are 
๎xor๎ed๎ (before the slopes are applied) at OUTPUT 1. The signal of 
OUTPUT 2 can be used in parallel. 
This allows you to for example to simulate single or repeated glitches. 
Command
:DIG:PATT:LOOP
Long
:DIGital[:STIMulus]:PATTern:LOOP[:LEVel[1]][:COUNt]
Form
Set & Query
Parameter
Numeric
*RST value
1
Specified Limits
1 to 2^20