User's Manual
5 Language Reference 
64  Series N5700 User’s Guide 
SCPI Command Summary 
NOTE 
Some [optional] commands have been included for clarity. All settings 
commands have a corresponding query. 
Subsystem Commands 
SCPI Command  Description 
ABORt  Aborts the triggered action 
CALibrate 
 :CURRent[:LEVel]  Calibrates the output current programming 
 :DATA <NRf>  Enters the calibration value 
 :DATE <”SPD”>  Sets the calibration date 
 :LEVel P1 | P2  Advances to the next calibration step 
 :PASSword <NRf>  Sets the numeric calibration password 
 :STATE <Bool> [,<NRf>]  Enables/disables calibration mode 
 :VOLTage[:LEVel]  Calibrates the output voltage programming 
INITiate 
 [:IMMediate][:TRANsient]  Initiates the trigger system 
 :CONTinuous[:TRANsient]  Enables/disables continuous triggers 
MEASure 
 [:SCALar]   
 :CURRent[:DC]?  Returns the measured output current 
 :VOLTage[:DC]?  Returns the measured output voltage 
OUTPut 
 [:STATe] <Bool>  Enables/disables the specified output 
 :PON   
 :STATe RST | AUTO  Programs the Power-On State 
 :PROTection   
 :CLEar  Resets latched protection 
[SOURce:] 
 CURRent   
 [:LEVel]   
  [:IMMediate][:AMPLitude] <NRf+>  Sets the output current 
  :TRIGgered[:AMPLitude] <NRf+>  Sets the triggered output current 
 :PROTection   
  :STATe <Bool>  Enables/disables over-current protection 
 VOLTage   
 [:LEVel]   
  [:IMMediate][:AMPLitude] <NRf+>  Sets the output voltage 
  :TRIGgered[:AMPLitude] <NRf+>  Sets the triggered output voltage 
 :LIMit   
  :LOW <NRf+>  Sets the low-voltage limit 
 :PROTection   
  [:LEVel] <NRf+>  Sets the over-voltage protection level 