User's Manual
Appendix B Verification and Calibration 
94  Series N5700 User’s Guide 
Verification 
Verification procedures verify that the power supply is operating 
normally and is within published specifications. There are two types 
of verification tests: 
Performance  These tests verify that the power supply meets all of the specifications listed 
in Appendix A. They can also be used to verify that the power supply is 
properly calibrated. 
Calibration  These procedures calibrate the power supply. 
If the power supply fails any of the tests or if abnormal test results 
are obtained, try calibrating the unit. If calibration is unsuccessful, 
return the unit to an Agilent Technologies repair facility (see 
Appendix D). 
Equipment Required 
The equipment listed in the following table, or the equivalent to this 
equipment, is required for the calibration and performance tests. A 
test record sheet may be found at the back of this section. 
Type Specifications  Recommended Model 
Digital Voltmeter  Resolution: 10 nV @ 1V; 
Readout: 8 1/2 digits 
Accuracy: 20 ppm 
Agilent 3458A or equivalent 
Current Monitor 
15 A (0.1Ω) 0.04%, TC=5ppm/°C 
100 A (0.01Ω) 0.04%, TC=5ppm/°C 
300 A (0.001Ω) 0.04%, TC=5ppm/°C 
Guildline 9230/15 
Guildline 9230/100 
Guildline 9230/300 
Load Resistor  For 750 W models: 
0.06Ω, 0.09Ω, 0.21Ω, 0.53Ω, 1.20Ω, 2.11Ω, 4.80Ω, 8.42Ω, 
13.33Ω, 30.0Ω, 120Ω, 462Ω - all resistors 1kW minimum. 
For 1500 W models: 
0.03Ω, 0.04Ω, 0.10Ω, 0.26Ω, 0.60Ω, 1.05Ω, 2.40Ω, 4.21Ω, 
6.67Ω, 15.0Ω, 60Ω, 231Ω - all resistors 2kW minimum. 
Electronic Load  150 V, 100 A minimum (for Models N5741- N5750A) 
150 V, 180 A minimum (for Models N5761- N5770A) 
Agilent N3300A mainframe, 
with 3 - N3305A modules 
GPIB Controller  Full GPIB capabilities (for calibrating the unit over the GPIB)  Agilent 82350B or 
equivalent 
Oscilloscope  Sensitivity: 1 mV 
Bandwidth Limit: 20 MHz 
Probe: 1:1 with RF tip 
Agilent Infiniium or 
equivalent 
RMS Voltmeter   True RMS 
 Bandwidth: 20 MHz 
 Sensitivity: 100 µV 
Rhode and Schwartz Model 
URE3 or equivalent 
Differential Amplifier  Bandwidth: 20 MHz  LaCroy 1855A, DA1850A, 
or equivalent 
Terminations 
1 – 50Ω BNC termination 
2 – 50Ω, 1/8 W termination resistors 
Variable-voltage xfmr 
or AC source 
Adjustable to highest rated input voltage range. 
Power: 2000 VA 
Agilent 6813B or equivalent 