User's Manual
6 Programming Examples 
84  Series N5700 User’s Guide 
Trigger Programming Example 
This example illustrates how to set up and trigger a voltage and 
current change. The voltage is measured before and after the trigger. 
Sub main_Trig() 
 Dim IDN As String 
 Dim IOaddress As String 
 Dim ErrString As String 
 Dim msg1 As String 
 ' This variable is used to monitor the status 
 Dim stat As Long 
 ' This variable controls the voltage 
 Dim VoltSetting As Double 
 ' This variable measures the voltage 
 Dim MeasureVolt As Double 
 ' This variable controls the current 
 Dim CurrSetting As Double 
 ' This variable represents the trigger current setting 
 Dim trigCurrSetting As Double 
 ' This variable controls the triggered voltage setting 
 Dim trigVoltSetting As Double 
 ' This constant represents the register value for Waiting for Trigger 
 Const WTG = 32 
 ' These variables are necessary to initialize the VISA COM 
 Dim ioMgr As AgilentRMLib.SRMCls 
 Dim Instrument As VisaComLib.FormattedIO488 
 ' The following line provides the VISA name of the GPIB interface 
 IOaddress = "GPIB0::5::INSTR" 
 ' Use the following line instead for LAN communication 
 ' IOaddress="TCPIP0::141.25.36.214" 
 ' Use the following line instead for USB communication 
 ' IOaddress = "USB0::2391::1799::US00000002" 
 ' Initialize the VISA COM communication 
 Set ioMgr = New AgilentRMLib.SRMCls 
 Set Instrument = New VisaComLib.FormattedIO488 
 Set Instrument.IO = ioMgr.Open(IOaddress) 
 VoltSetting = 3 ' volts 
 CurrSetting = 2 ' amps 
 trigVoltSetting = 5 ' volts 
 trigCurrSetting = 3 ' amps 
 With Instrument 
 ' Send a power reset to the instrument 
 .WriteString "*RST" 
 ' Query the instrument for the IDN string 
 .WriteString "*IDN?" 
 IDN = .ReadString 