4 Operating the Power Supply Remotely 
62  Series N5700 User’s Guide 
SCPI Command Completion 
SCPI commands sent to the power supply are processed either 
sequentially or in parallel. Sequential commands finish execution 
before a subsequent command begins. Parallel commands allow other 
commands to begin executing while the parallel command is still 
executing. 
The following is a list of parallel commands. You should use some 
form of command synchronization as discussed in this section before 
assuming that these commands have completed. 
OUTPut:STATe INITiate 
VOLTage OUTPut:PROTection:CLEar 
CURRent 
The *WAI, *OPC, and *OPC? common commands provide different 
ways of indicating when all transmitted commands, including any 
parallel ones, have completed their operations. Some practical 
considerations for using these commands are as follows: 
*WAI
This command prevents the power supply from processing subsequent 
commands until all pending operations are completed. For example, 
the *WAI command can be used to make a voltage measurement after 
an output on command has completed:
OUTPut ON;*WAI;:MEASure:VOLTage?
*OPC?
This command places a 1 in the Output Queue when all pending 
operations have completed. Because it requires your program to read 
the returned value before executing the next program statement, 
*OPC? can be used to cause the controller to wait for commands to 
complete before proceeding with its program. 
*OPC
This command sets the OPC status bit when all pending operations 
have completed. Since your program can read this status bit on an 
interrupt basis, *OPC allows subsequent commands to be executed. 
NOTE 
The trigger subsystem must be in the Idle state for the status OPC bit to be 
true. As far as triggers are concerned, OPC is false whenever the trigger 
subsystem is in the Initiated state. 
Device Clear 
You can send a Device Clear at any time to abort a SCPI command 
that may be hanging up the GPIB interface. Device Clear clears the 
input and output buffers of the power supply. The status registers, 
error queue, and all configuration states are left unchanged by Device 
Clear. Device Clear also prepares the power supply to accept a new 
command string. The following statement shows how to send a device 
clear over the GPIB interface using Agilent BASIC: 
CLEAR 705    IEEE-488 Device Clear 