4 Operating the Power Supply Remotely 
56  Series N5700 User’s Guide 
Factory-shipped LAN Settings 
The factory-shipped LAN settings documented in the following table 
are optimized for connecting your power supply to a site network. 
They should also work well for other network configurations. 
The factory-shipped settings can be restored by pressing and holding 
the front panel LAN button for three seconds. Pressing the LAN 
button again while the message “LAn rES” is displayed resets the 
LAN settings. 
Factory-shipped non-volatile LAN settings 
Get IP Address  Automatic  Dynamic DNS naming service  Enabled 
IP Address  169.254.57.0  NetBIOS naming service  Enabled 
Subnet Mask  255.255.0.0  Domain name  Blank 
Default Gateway  0.0.0.0  TCP keepalive  Enabled 
Obtain DNS server from DHCP  Enabled  TCP keepalive seconds  1800 
DNS server  Blank  Ethernet Auto-negotiation  Enabled 
Host name  A-N57xxA-xxxxx  Ping server  Enabled 
 Web password Blank 
Using the Setup Utility 
A Setup utility that lets you configure the LAN settings of your 
instrument is provided on the Product Reference CD-ROM included 
with this manual. Install and run this Setup utility if you cannot 
configure the LAN interface as previously described in this section. 
1  Connect your power supply to your computer using either the 
USB interface or the GPIB interface as previously described. 
2  Install the Setup utility on your computer. Run the Setup utility 
by clicking Start|Programs|Agilent|N5700 Setup Utility. 
3  Configure the following LAN address parameters. These are 
located under the Settings tab. For a description of these 
parameters, refer to the previous section. 