Service requests are enabled for control sockets using the Service 
Request Enable register. Once service requests have been enabled, 
the client program listens on the control connection. When SRQ goes 
true the instrument will send the string “SRQ +nn” to the client. The 
“nn” is the status byte value, which the client can use to determine 
the source of the service request. 
Configuring the LAN Parameters 
To configure the LAN parameters from the instrument’s Web server, 
launch the Web server as previously described, and click on the View 
& Modify Configuration tab on the left side of the page. Then click on 
the Modify Configuration button on the top of the page. The following 
screen lets you modify the LAN parameters: 
The configurable LAN parameters are described as follows: 
Obtain IP 
Address 
This parameter configures the addressing of the instrument. Auto 
automatically configures the addressing. When selected, the instrument first 
tries to obtain an IP address from a DHCP server. If a DHCP server is found, 
the DHCP server assigns an IP address, Subnet Mask, and Default Gateway to 
the instrument. If a DHCP server is unavailable, the instrument tries to obtain 
an IP address using AutoIP. AutoIP automatically assigns an IP address, 
Subnet Mask, and Default Gateway addresses on networks that do not have a 
DHCP server. Manual allows you to manually configure the addressing of the 
instrument by entering values in the following three fields. 
IP Address  This value is the Internet Protocol (IP) address of the instrument. An IP 
address is required for all IP and TCP/IP communications with the instrument. 
An IP Address consists of 4 decimal numbers separated by periods. Each 
decimal number ranges from 0 through 255. 