User's Manual
Operating the Power Supply Remotely 4 
Series N5700 User’s Guide  53 
Using Telnet 
In an MS-DOS Command Prompt box type: telnet hostname 5024 
where hostname is the N5700 hostname or IP address, and 5024 is 
the instrument’s telnet port. 
You should get a Telnet session box with a title indicating that you 
are connected to the power supply. Type the SCPI commands at the 
prompt. 
Using Sockets 
Agilent instruments have standardized on using port 5025 for SCPI 
socket services. A data socket on this port can be used to send and 
receive ASCII/SCPI commands, queries, and query responses. All 
commands must be terminated with a newline for the message to be 
parsed. All query responses will also be terminated with a newline. 
The power supply allows any combination of up to three 
simultaneous data socket and telnet connections to be made. 
The socket programming interface also allows a control socket 
connection. The control socket can be used by a client to send device 
clear and to receive service requests. Unlike the data socket, which 
uses a fixed port number, the port number for a control socket varies 
and must be obtained by sending the following SCPI query to the data 
socket: SYSTem:COMMunicate:TCPip:CONTrol? 
After the control port number is obtained, a control socket 
connection can be opened. As with the data socket, all commands to 
the control socket must be terminated with a newline. All query 
responses will also be terminated with a newline. 
To send a device clear, send the string “DCL” to the control socket. 
When the power supply has finished performing the device clear it 
echoes the string “DCL” back to the control socket. 