User's Manual
Appendix D Compatibility 
124  Series N5700 User’s Guide 
Differences – In General 
The following table documents the general differences between the 
way Compatibility commands work on the Agilent N5700 power 
supplies and the way they worked on the Agilent 603xA power 
supplies. 
Item Differences 
Queries  The Agilent N5700 will respond to multiple queries. 
It will not allow a space separator between numbers. 
It will not allow a user to query information, read back only a portion of the information, 
send another command, and finish reading back the information from the original query. 
Sending a second query without reading the response to the first will generate an error. 
Model number queries will only return the N5700 model numbers. 
Status functions  Serial Poll will be controlled by the SCPI status model and will not act like a 603xA 
power supply. 
SRQ will be controlled by the SCPI status model. 
Parallel poll will not work. 
Settings  The full-scale limits will match the Agilent N5700 limits. 
Measurement  Floating point numbers returned by the instrument may not have exactly the same 
syntax or number of digits. 
Calibration  Calibration must be done in SCPI. 
Storage states  The Agilent N5700 units have 16 volatile states. 