User's Manual
Appendix B Verification and Calibration 
114  Series N5700 User’s Guide 
Step 5. Select the first voltage calibration point. 
CAL:LEV P1 
*OPC? 
Step 6. Measure the output voltage and enter the data. 
CAL:DATA <data> 
Step 7. Select the second voltage calibration point. 
CAL:LEV P2 
*OPC? 
Step 8. Measure the output voltage and enter the data. 
CAL:DATA <data> 
Step 9. Exit calibration mode. 
CAL:STAT OFF 
Current Programming and Measurement Calibration 
Step 1. Connect a precision shunt resistor to an output. Connect the Agilent 
3458A across the shunt. The shunt should be able to measure at least 
120% of the power supply’s rated full-scale current. 
Step 2. Enable current calibration mode. 
*RST 
OUTP ON 
CAL:STAT ON 
Step 3. Set the output voltage high enough to compensate for any voltage 
drops on the load leads and current shunt. 
VSET 0.5 
Step 4. Select current calibration. 
CAL:CURR 
Step 5. Select the first current calibration point. 
CAL:LEV P1 
*OPC? 
Step 6. Calculate the shunt current (I=V/R) and enter the data. 
CAL:DATA <data> 
Step 7. Select the second current calibration point. 
CAL:LEV P2 
*OPC? 
Step 8. Calculate the shunt current (I=V/R) and enter the data. 
CAL:DATA <data> 
Step 9. Exit calibration mode. 
CAL:STAT OFF 