User's Manual
Verification and Calibration Appendix B 
Series N5700 User’s Guide  99 
Constant Current Tests 
NOTE 
Refer to the appropriate test record form for the instrument settings of the 
model you are checking. 
Current Programming and Readback Accuracy 
Test category = performance, calibration 
This test verifies that the current programming and measurement 
functions are within specifications. 
1  Turn off the power supply and connect the current shunt directly 
across the output. Connect the DVM across the current shunt. 
2  Turn on the power supply and program the output voltage to its 
full-scale value and the output current to zero. The CC 
annunciator should be on and the output voltage reading should 
be approximately zero. 
3  Divide the voltage drop (DVM reading) across the current shunt 
by its resistance to convert to amps and record this value (Iout). 
Also record the current reading on the front panel display. The 
readings should be within the limits specified in the test record 
card for the appropriate model under Current Programming and 
Readback, Minimum Current Iout. 
4  Program the output current to its full-scale rating. 
5  Divide the voltage drop (DVM reading) across the current shunt 
by its resistance to convert to amps and record this value (Iout). 
Also record the current reading on the front panel display. The 
readings should be within the limits specified in the test record 
card for the appropriate model under Current Programming and 
Readback, High Current Iout. 
CC Load Effect 
Test category = performance 
This test measures the change in output current resulting from a 
change in output voltage from full scale to short circuit. 
1  Turn off the power supply and connect the current shunt, DVM, 
and electronic load as shown in figure B. Connect the DVM 
directly across the current shunt. 
2  To ensure that the values read during this test are not the 
instantaneous measurement of the AC peaks of the output 
current ripple, several DC measurements should be made and 
averaged. If you are using an Agilent 3458A, you can set up the 
voltmeter to do this automatically. From the instrument’s front 
panel, program 100 power line cycles per measurement. Press 
NPLC 100 ENTER. 
3  Turn on the power supply and program the output current to its 
full-scale value and the output voltage to its maximum 
programmable value (Vmax). 