User's Manual
Installation 2 
Series N5700 User’s Guide  25 
The J2 connector plug specifications are as follows: 
Plug Type:
  MC 1.5/5-ST-3.81, Phoenix 
Wire Size:
  AWG 28 to AWG 16 
Stripping Length:
  7 mm (0.28 in.) 
Torque:
  0.22 – 0.25 Nm (1.95 – 2.21 in-lb.) 
Local Sensing 
The power supply is shipped with the rear panel J2 sense connector 
wired for local sensing of the output voltage. With local sensing, the 
output voltage regulation is made at the output terminals. This 
method does not compensate for voltage drop on the load wires, 
therefore it is recommended only for low load current applications or 
where the load regulation is less critical. The following figure 
illustrates the internal connections of the J2 connector. 
NOTE 
If the power supply is operated without the local sense jumpers or without the 
remote sense lines connected, it will continue to work, but the output voltage 
regulation will be degraded. Also, the OVP circuit may activate and shut down 
the power supply. Note that the internal wiring between +V and + local sense 
and between –V and – local sense will fail if load current flows through it. 
Remote Sensing 
Use remote sensing in applications where load regulation at the load 
is critical. Remote sensing allows the power supply to automatically 
compensate for the voltage drop in the load leads. Refer to Appendix 
A for the maximum allowable voltage drop on the load wires. 
Remote sensing is especially useful in constant voltage mode with 
load impedances that vary or have significant lead resistance. It has 
no effect in constant current mode. Because sensing is independent 
of other power supply functions it can be used regardless of how the 
power supply is programmed. With remote sensing, voltage readback 
monitors the load voltage at the remote sense points. 
Use twisted or shielded wires to minimize noise pick-up. If shielded 
wires are used, the shield should be connected to the ground at one 
point, either at the power supply chassis or the load ground. The 
optimal point for the shield ground should be determined by 
experimentation. 
-
-
Rem.sense
Local sense
Local sense
Rem.sense
+
Load lines, twisted
pair, shortest length
possible.
+V
-V
Load
+
Power
Supply
+
Error
Amp.