118  Series N5700 User’s Guide 
Error Messages 
Displaying the SCPI error queue 
The entire error queue is read, then emptied, using the following 
command: SYST:ERR? 
Error List 
The following table documents the various error messages that the 
power supply supports: 
Error Device-dependent Errors (these errors set Standard Event Status register bit #3) 
0  No error 
This is the response to the ERR? query when there are no errors. 
100  Too many channels 
You have specified more channels than are installed in the mainframe. 
101  Calibration state is off 
Calibration is not enabled. The instrument will not accept calibration commands. 
102  Calibration password is incorrect 
The calibration password is incorrect. 
104  Bad sequence of calibration commands 
Calibration commands have not been entered in the proper sequence. 
114  CAL:DATE must be yyyy/mm/dd 
The calibration date must be entered in the numeric format yyyy=year, mm=month, dd=date 
203  Compatibility function not implemented 
The requested compatibility function is not available. 
204  NVRAM checksum error 
A checksum error has occurred in the instrument’s nonvolatile random access memory. 
205  NVRAM full 
The nonvolatile random access memory of the instrument is full. 
206  File not found 
The internal calibration file or the internal channel attribute file was not found in NVRAM. 
209  Output communications failure 
A hardware failure has occurred on the power supply. 
302  Option not installed 
The option that is programmed by this command is not installed. 
351  VOLT setting conflicts with VOLT:PROT setting 
Attempted to program the voltage above the over-voltage protection setting. 
352  VOLT:PROT setting conflicts with VOLT setting 
Attempted to set the over-voltage protection below the voltage setting. 
353  VOLT setting conflicts with VOLT:LIM:LOW setting 
Attempted to program the voltage below the under-voltage limit setting. 
354  VOLT:LIM:LOW setting conflicts with VOLT setting 
Attempted to set the under-voltage limit above the voltage setting 