User's Manual
Aborting a Calibration in Progress 
Sometimes it may be necessary to abort a calibration after the procedure
has already been initiated. You can abort a calibration at any time by
pressing any front-panel key (except
 Shift-Cancel  ). When performing a
calibration from the remote interface, you can abort a calibration by
issuing a remote interface device clear message or by pressing the
front-panel 
LOCAL key. 
Frequency and Burst Rate Adjustment 
The function generator stores two calibration constants related to
frequency and burst rate output. The constants are calculated from the
adjustment value entered and are stored at the completion of each setup.
1 Use a frequency meter to measure the function generator output
frequency for each setup in the following table. These adjustments use 
a 50
W output termination.
Nominal Output
SETUP FREQUENCY AMPLITUDE
 00 * 1.00 kHz 10 Vpp
Adjustment for main frequency
generator, sine wave output
01 500 Hz 10 Vpp
Adjustment for burst rate timing,
pulse output.
2 Use the CALIBRATE menu to adjust the displayed frequency at each
setup to match the measured frequency and enter the value. 
3 Perform the Frequency Verification procedures on page 56.
* A new calibration (SETUP 86 – Rev 4.0) has been added as an alternative to SETUP 00.
 The new calibration outputs a 10 MHz sine wave, rather than the 1 kHz signal used for
 SETUP 00. The new calibration reduces slew rate dependent errors in the frequency
 measurement and is especially important when calibrating the Phase-Lock Assembly
 (Option 001). Note that either setup is sufficient to calibrate the carrier frequency and
 you don’t need to perform both.
 4
Chapter 4 Calibration Procedures
Aborting a Calibration in Progress 
69










