User's Manual
 <    >  7 Move the flashing cursor over the digit to be edited. 
 1
 ^ 
     8 Change the value in the display to match the measured frequency. 
 1.000,0040KHz
 Enter  9 Calculate and save the new value. 
 CALIBRATING
10 Perform the next adjustment procedure.
The setup number and function generator output is automatically set for
the next adjustment procedure.
 SETUP ^01
You will press ENTER twice for each calibration step, once to select 
the setup (as described in step 6) and once to enter the adjustment 
(as described in step 9).
1
You can also use the knob to scroll left or right through the choices on each level of
the menu
Chapter 4 Calibration Procedures
General Calibration/Adjustment Procedure 
68