User's Manual
To unsecure the function generator for calibration
The function generator can use a calibration security code to prevent
unauthorized or accidental calibration. This procedure shows you how to
unsecure the function generator for calibration. 
 Shift  1 Turn on the menu.
 Menu On/Off   A: MOD MENU 
 <   2 Move across to the CAL MENU choice on this level. 
 F: CAL MENU
     3 Move down a level to the SECURED command.
 1: SECURED
If the display shows UNSECURED, you do not need to perform this
procedure to execute a calibration.
 3
Chapter 3 Front-Panel Menu Operation
To unsecure the function generator for calibration 