User's Manual
 549
Text Only GPIB Errors 
Un-numbered (text only) GPIB error messages are generally self-explanatory. For 
example, trying to retrieve a saved register that does not exist generates the following 
error message: 
HP-IB Error: Register does not exist. 
The following list contains a subset of the Test Set’s text only GPIB error messages. These 
messages represent error conditions which may require explanation in addition to the error 
message text. 
HP-IB Error during Procedure catalog. Check Config.
This error occurs when the Test Set fails to access an external GPIB disk drive when 
trying to obtain a catalog of procedure files. This would occur when the Select 
Procedure Location: field on the TESTS (Main Menu) screen is set to Disk and 
the operator then tries to select a procedure filename using the Select Procedure 
FIlename: field. Ensure that the Mode field on the I/O CONFIGURE screen is set to 
Control and that the External Disk Specification field on the TESTS 
(External Devices) screen has the correct mass storage volume specifier for the external 
disk drive. 
HP-IB Query Error. Check instrument state.
This message usually appears when the control program queries a measurement that is not 
currently available (on the currently displayed screen and in the ON state), such as 
querying the TX Frequency measurement when TX Freq Error is displayed. 
This message may also be immediately followed by the message, 
HP-IB Error: 
−