User's Manual
304
S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb  
Chapter 5, Advanced Operations
Instrument Initialization
Methods of Initialization 
There are six methods of initializing the Test Set: 
Power On Reset 
Front panel PRESET key 
*RST IEEE 488.2 Common Command 
Device Clear (DCL) GPIB Bus Command 
Selected Device Clear (SDC) GPIB Bus Command 
Interface Clear (IFC) GPIB Bus Command 
When the Test Set is initialized some fields are “restored” (put back to their 
default state), some fields are “maintained” (kept at their current state or value), 
and some fields are “initialized” (returned to their default value). 
The following sections discuss the effects each of the six initialization methods 
has on the Test Set. 
Power-On Reset 
The Power-On Reset is accomplished by applying or cycling AC/DC power to the 
Test Set. 
For the CONFIGURE, PRINT CONFIGURE, TESTS (Execution Conditions), 
TESTS (Printer Setup) and I/O CONFIGURE screens, 
Table 28 lists the fields 
which are restored/initialized when the Test Set AC/DC power is cycled. The 
restored state or initialized value is listed below the field name. Fields which are 
not listed are maintained at their current value, whatever that may happen to be. 
All fields in the TESTS (Main Menu) screen and the TESTS (External Devices) 
screen are maintained at their current state/value. The current state/value of the 
maintained fields can be ascertained programmatically. 