User's Manual
 29
Chapter 1, Using GPIB
Overview of the Test Set
Differences Between the Test Set’s IBASIC Controller and Other Single-Tasking 
Controllers
The IBASIC Controller is unlike other single tasking instrumentation controllers 
in several ways. First, it does not have a keyboard. This imposes some limitations 
on creating and editing IBASIC programs directly on the Test Set. In Internal 
Automatic Control mode a “virtual” keyboard is available in firmware which 
allows the operator to enter alphanumeric data into a dedicated input field using 
the rotary knob. This is not the recommended programming mode for the IBASIC 
Controller. This feature is provided to allow user access to IBASIC programs for 
short edits or troubleshooting. Several programming modes for developing 
IBASIC programs to run on the internal IBASIC Controller are discussed in this 
manual.
Secondly, the IBASIC Controller has a dedicated GPIB interface, select code 8 in 
Figure 2, for communicating with the internal instruments of the Test Set. This 
GPIB interface is only available to the IBASIC Controller. There is no external 
connector for this GPIB interface. No external instruments may be added to this 
GPIB interface. The GPIB interface, select code 7 in 
Figure 2, is used to interface 
the Test Set to external instruments or to an external controller. The dedicated 
GPIB interface at select code 8 conforms to the IEEE 488.2 Standard in all 
respects but one. The difference being that each instrument on the bus does not 
have a unique address. The Instrument Control Hardware determines which 
instrument is being addressed through the command syntax. Refer to 
Chapter 4, 
“GPIB Commands,”
 for a listing of the GPIB command syntax for the Test Set.










