User's Manual
 183
Equivalent Front-Panel Key Commands
To Turn Measurement Averaging ON and OFF. Use the :AVERage:STATe 
commands to turn the averaging data function ON and OFF. 
Syntax
:AVER age:STATe ON
:AVERage:STATe OFF
Example
OUTPUT 714;"MEAS:AFR:DISTN:AVER:STAT ON" 
This turns the AVG Data Function ON for the Audio Frequency 
Analyzer Distortion measurement. 
To Query the Measurement Averaging State. Use the :AVERage:STAT? 
commands to query the current state of the averaging data function. The returned 
value is either: 0 (OFF) or 1 (ON). 
Syntax
:AVERage:STAT?
Example
OUTPUT 714;"MEAS:AFR:DISTN:AVER:STAT?”
ENTER 714;State_on_off ! 1 = ON, 0 = OFF 
This queries the state of the AVG Data Function for the Audio 
Frequency Analyzer Distortion measurement. 