Equivalent Front-Panel Key Commands
DATA FUNCTIONS Keys 
The Data Functions keys can be divided into two groups; those which affect 
measurements (REF SET, METER, AVG, HI LIMIT and LO LIMIT), and those 
which affect numeric entry fields (INCR
÷10, INCR SET, INCR×10, Up-arrow, 
Down-arrow). For measurements, the Data Functions enable the programmer to 
change the way measurements are calculated and displayed, and provide 
measurement limit detection. For numeric entry fields, the Data Functions enable 
the programmer to set, scale, and change the field’s increment value. Each Data 
Function is described in detail in the Test Set’s User’s Guide. 
Refer to the 
“Number Measurement Syntax” on page 177 for full command syntax. 
Guidelines for Using Measurement Data Functions 
• Data Functions are turned ON and OFF for individual measurements. The GPIB Data 
Function commands must immediately follow the GPIB command for the individual 
measurement. For example, to turn the AVG Data Function ON for the Audio 
Frequency Analyzer Distortion measurement, the following command string would be 
sent to the Test Set: 
OUTPUT 714;"MEAS:AFR:DISTN:AVER:STAT ON"
• Attribute Units (AUNits) are used with the Data Functions to specify the units-of-
measure for numeric data which is read or set through GPIB. Refer to “Attribute Units 
(AUNits)” on page 81. 
• Data Function settings, such as Number of Averages or Reference value, are retained if 
the function is turned off. The setting values are initialized or changed under the 
following conditions: 
• The Test Set is turned off. 
• The Test Set is PRESET. 
• A saved register is recalled. 