User's Manual
DL Display Line
DL
Display Line
Defines the level of the display line in the active amplitude units and displays the display line
on the spectrum analyzer screen.
Syntax
Item
Description/Default
Range
Number
Any real or integer number. Default units are dBm.
Dependent on the
reference level.
Equivalent
Softkey:
l3SP
LINE
OM
OFF .
Preset State: DL OFF
Step Increment: 1 major division.
Related Commands:
AUNITS,
AUTO, LG, LN, NRL, RL, ROFFSET, TH.
Example
OUTPUT 718;"AUNITS
DBM;"
Changes the active amplitude units to
dBm.
OUTPUT
718;"DL
ON;"
Turns on the
displa?/
line.
OUTPUT
718;"DL
-5DM;"
Changes display line to -5
dBm.
Description
Activating video trigger mode activates the display line. The AUTO command and “DL OFF;
”
turn off the display line. See
“AUNITS”
for more information on changing the active amplitude
units.
Programming Commands 5-135