User's Manual
188
Chapter 5: Reference
The Trigger Tab
Sets up to never trigger. You must select the stop button to view the
captured data. For maximum data capture the trigger position should be
set to end.
A warning message concerning no trigger action will be displayed in the
Run Status window. This warning message can be ignored since the intent
of the trigger function is not to trigger. The run status window can be
prevented from coming up with this warning by setting the Popup On
Severity level to errors only.
Settings Subtab
Sample Period (Timing mode only). Lets you specify how often the logic
analyzer samples signals from the device under test. In the
timing mode's 800 MHz / 64M Sample Conventional
configuration, the sample period is fixed at 1.25 ns.
Acquisition
Depth Lets you use a smaller portion of the acquisition memory
and speed up processing of the captured data.
Trigger Position Lets you specify where the sample that triggered the
analyzer should appear among all the other samples that
are stored in acquisition memory.
Count Lets you turn off counting, count time for each stored
sample, or, in the state mode's 200 Mb/s / 32M State
configuration, count the occurrence of some state with
each stored sample.
Intermodule
Control Opens the Intermodule window which lets you configure
multiple instrument measurements, adjust the order of