User's Manual

3-5
In Case of Difficulty
If the Plug-in Does Not Operate
After viewing the signal, press
triggered
.
Make sure Channel Display is on by pressing:
Channel
,
Display on off
Make sure the channel offset is adjusted so the waveform is not clipped off
the display.
If you are using the plug-in module only as a trigger source, make sure at
least one other channel is turned on.
If all of the channels are turned off, the mainframe will
not
trigger.
Make sure the mainframe identifies the plug-in module by pressing:
Utility
,
System config...
The calibration status of the plug-in modules is listed near the bottom of the
display, in the box labeled "
Plug-ins
". If the model number of the plug-in
module is listed next to the appropriate slot number, then the mainframe
has identified the plug-in.
If "
~known"
is displayed, instead of the model number of the plug-in
module, remove and reinsert the plug-in module in the same slot. If
"
~known
" is still displayed, then the memory contents of the plug-in module
are corrupt. Refer to the service guide for the plug-in module or contact a
qualified service department. If all of the above steps check out okay, and
the plug-in module still does not operate properly, then the problem is
beyond the scope of this book. Refer to the plug-in module service guide or
return the plug-in to a qualified service department.