98 Agilent 1260 Infinity High Performance Autosampler User Manual
7 Error Information
Module Error Messages
Missing Vial or Missing Wash Vial
Error ID: 4019, 4034, 4035, 4541, 4542, 4706, 4707
No vial was found in the position defined in the method or sequence.
When the needle carrier moves to a vial and the needle goes into the vial,
the position of the needle is monitored by an encoder behind the vial
pusher. If no vial is present, the encoder detects an error and the message
“missing vial” is generated.
Probable cause Suggested actions
1 No vial in the position defined in the method
or sequence.
Install the sample vial in the correct position, or
edit the method or sequence accordingly.
2 Defective needle carrier assembly.
Please contact your Agilent service
representative.
3 Defective transport assembly flex board.
Please contact your Agilent service
representative.
4 Defective main board.
Please contact your Agilent service
representative.