Agilent 1260 Infinity High Performance Autosampler User Manual 97
Error Information
7
Module Error Messages
Needle to Needle Seat Position
Error ID: 4510, 4511, 4714
The needle failed to reach the end position in the needle seat.
The position of the needle is monitored by a position encoder on the needle
carrier. If the needle fails to reach the end point, or if the encoder fails to
recognize the needle carrier movement, the error message is generated.
Needle Carrier Failed
The needle carrier on the Sample Transport Assembly failed to move correctly.
Probable cause Suggested actions
1 Bad sample transport/sampling unit
alignment
Do an auto-alignment
2 Bent needle.
Check and exchange the needle assembly if
necessary.
3 Missing needle.
Exchange the needle carrier assembly.
4 Blocked seat.
Clean or change the needle seat assembly if
necessary.
5 Defective position sensor in the needle
carrier assembly.
Please contact your Agilent service representative.
6 Defective main board.
Please contact your Agilent service representative.
Probable cause Suggested actions
1 Defective Z-motor.
Please contact your Agilent service representative.
2 Vial pusher blocked.
Please contact your Agilent service representative.
3 Bad needle carrier positioning in X or Theta.
Please contact your Agilent service representative.
4 Defective vial pusher sensor.
Please contact your Agilent service representative.
5 Defective main board.
Please contact your Agilent service representative.