User manual

Agilent 1260 Infinity High Performance Autosampler User Manual 95
Error Information
7
Module Error Messages
Valve to Bypass Failed
Error ID: 4014, 4701
The injection valve failed to switch to the bypass position.
The switching of the injection valve is monitored by two microswitches on
the valve assembly. The switches detect the successful completion of the
valve movement. If the valve fails to reach the bypass position, or if the
microswitch does not close, the error message is generated.
Valve to Mainpass Failed
Error ID: 4015
The injection valve failed to switch to the mainpass position.
The switching of the injection valve is monitored by two microswitches on
the valve assembly. The switches detect the successful completion of the
valve movement. If the valve fails to reach the mainpass position, or if the
microswitch does not close, the error message is generated.
Probable cause Suggested actions
1 Valve in an intermediate position between
the bypass and mainpass positions.
Turn the Autosampler main power OFF and ON.
2 Defective injection valve.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Valve in an intermediate position between
the bypass and mainpass positions.
Turn the Autosampler main power OFF and ON.
2 Defective injection valve.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.