How to Achieve Lowest Carry Over
How to Achieve Lowest Carry Over
Carryover is measured when residual peaks from a previous
active- containing injection appear in a subsequent blank solvent injection.
There will be carry over between active injections which may lead to
erroneous results. The level of carryover is reported as the area of the
peak in the blank solution expressed as a percentage of the area in the
previous active injection. The Agilent 1260 Infinity Autosampler is
optimized for lowest carryover by careful design of the flow path and use
of materials in which sample adsorption is minimized. A carryover figure
of 0.002 % should be achievable even when a triple quadrupole mass
spectrometer is the detector. Operating settings of the autosampler allow
the user to set appropriate parameters to minimize carryover in any
application involving compounds liable to stick in the system.
The following functions of the autosampler can be used to minimize
carryover:
• Internal needle wash
• External needle wash
• Needle seat backflush
• Injection valve cleaning
The flow path, including the inside of the needle, is continuously flushed
in normal operation, providing good elimination of carryover for most
situations. Automated delay volume reduction (ADVR) will reduce the
delay volume but will also reduce the flushing of the autosampler and
should not be used with analytes where carryover might be a problem.
The outside of the needle can be washed using a wash vial in a specific
location or the needle can be washed using the flush port. If a wash vial
in a tray location specified by the user is chosen then this vial should
have no septum and should contain a solvent suitable for washing the
sample from the needle. The septum is not used to avoid wiping
contamination off the needle on the downstream only to re- apply it on the
upstroke. The needle can be dipped into the vial multiple times. This will
be effective in removing a small degree of carryover but for more effective
washing of the outside of the needle use the flushport.