User manual

76 Agilent 1260 Infinity High Performance Autosampler User Manual
5 Optimizing Performance
How to Achieve Higher Sensitivity
How to Achieve Higher Sensitivity
The sensitivity of a separation method is linked to the choice of stationary
and mobile phases as good separation with narrow peaks and a stable
baseline with minimal noise are desirable. The choice of instrument
configuration will have an effect and a major impact is the setup of the
detector. This section considers how sensitivity is affected by:
Pump mixer volume
Narrower columns
Detector flow cell
Detector parameters
In addition, the discussion on detector parameters also mentions the
related topics of selectivity and linearity.
Columns
Sensitivity is specified as a signal- to- noise ratio (S/N) and hence the need
to maximize peak height and minimize baseline noise. Any reduction in
peak dispersion will help to maintain peak height and so extra- column
volume should be minimized by use of short, narrow internal diameter,
connection capillaries and correctly installed fittings. Using smaller inner
diameter columns should result in higher peak height and is therefore
ideal for applications with limited sample amounts. If the same sample
amount can be injected on a smaller i.d. column, then the dilution due to
column diameter will be less and the sensitivity will increase. For
example, decreasing the column i.d. from 4.6 mm to 2.1 mm results in a
theoretical gain in peak height of 4.7 times due to the decreased dilution
in the column. For a mass spectrometer detector, the lower flow rates of
narrow columns can result in higher ionization efficiencies and therefore
higher sensitivity.