5 Optimizing Performance
How to Achieve High Throughput
How to Achieve High Throughput
The injection can be optimized for speed remembering that drawing the
sample too fast can reduce the reproducibility. Marginal gains are to be
made here as the sample volumes used tend towards the smaller end of
the range in any case. A significant portion of the injection time is the
time taken with the needle movements to and from the vial and into the
flush port. These manipulations can be performed while the previous
separation is running. This is known as "overlapped injection" and it can
be easily turned on from the autosampler setup screen in the control
software. The autosampler can be told to switch the flow through the
autosampler to bypass after the injection has been made and then after,
for example, 3 minutes into a 4 minutes run to start the process of
aspirating the next sample and preparing for injection. This can typically
save 0.5 to 1 minute per injection.